Marshall Stability Test Apparatus
- Product Type:Marshall Stability Test Apparatus
- Primary Application:Determination of Marshall Stability and Flow of Bituminous Mixtures
The Softening Point Apparatus is an automatic laboratory testing instrument designed to determine the softening point of bitumen and other bituminous materials using the Ring and Ball method.
The softening point is defined as the temperature at which a material attains a specified degree of softening under controlled test conditions. The apparatus provides an automatic temperature rise of 5°C ± 0.5°C per minute and features a digital temperature indicator with a least count of 0.1°C for precise temperature monitoring.
The system includes brass tapered rings, steel balls, ball-centering guides, a regulated electrical stirrer, immersion heater, RTD probe, and bottom-mounted magnetic stirrer. It is designed to meet the requirements of IS 1205:1958 for softening point determination.
Bulk pricing available
| Technical Specifications | |
|---|---|
| Product Type | Automatic Softening Point Apparatus |
| Testing Method | Ring and Ball Method |
| Primary Application | Determination of Softening Point of Bitumen |
| Temperature Range | Up to 135°C |
| Temperature Rise Rate | 5°C ± 0.5°C per minute |
| Digital Temperature Indicator | Provided |
| Temperature Indicator Least Count | 0.1°C |
| Steel Ball Diameter | 9.5 mm |
| Number of Steel Balls | 2 |
| Ring Material | Brass |
| Number of Tapered Rings | 2 |
| Ring Depth | 6.4 mm |
| Ring ID at Top | 17.5 mm |
| Ring ID at Bottom | 15.9 mm |
| Ring OD | 20.6 mm |
| Ball Centering Guides | 2 |
| Ring Holder | 1 |
| Beaker | Borosil Glass |
| Beaker Internal Diameter | Approx. 8.5 cm |
| Beaker Depth | Approx. 12 cm |
| Stirrer | Regulated Electrical Stirrer |
| Stirrer Shaft | Adjustable |
| Heating System | Immersion Heater with Controller |
| Temperature Sensor | RTD Probe |
| Magnetic Stirrer | Bottom-Mounted |
| Applicable Standard | IS 1205:1958 |




Essential technical, operational, and standard compliance answers for the Softening Point Apparatus.
The Softening Point Apparatus is used to determine the softening point of bitumen and other bituminous materials under specified test conditions using the Ring and Ball method.
The Ring and Ball method determines the temperature at which a bituminous material softens sufficiently for a steel ball placed on the sample to move through a specified distance under controlled heating conditions.
The apparatus provides a temperature range of up to 135°C.
The system provides an automatic temperature rise of 5°C ± 0.5°C per minute.
The apparatus includes two tapered brass rings with a depth of 6.4 mm, top internal diameter of 17.5 mm, bottom internal diameter of 15.9 mm, and external diameter of 20.6 mm.
The apparatus is designed to conform to the requirements of IS 1205:1958 for determining the softening point of bituminous materials.
We stock tripods, leveling staffs, and protective cases for all our surveying instruments.