Office B-814, WTT Bh. Skoda showroom, SG Highway, Makarba, Ahmedabad GST 24BVRPK5410M1Z9
TrustSEAL Verified marutisagar@yahoo.in
Cement Testing Instruments

Laboratory Cement Autoclave

Highly Accurate Premium Quality Warranty Included

The Laboratory Cement Autoclave is a high-performance instrument engineered by Maruti Sagar Engineering for accurate quality control, material testing evaluation, and standard compliance.

The Laboratory Cement Autoclave is a high-quality cement testing instruments designed for civil engineering laboratories and field testing compliance.

Bulk pricing available

Request a Quote

Get Quote
Technical Specifications
Chamber Material Heavy-duty Stainless Steel (SS 304)
Operating Pressure Up to 21 kg/cm² (300 psi)
Temperature Range Up to 215°C
Safety Controls Pressure switch & safety valves
Compliance IS 4031 (Part 3) / ASTM C151
FAQs

Frequently Asked Questions

Essential technical, operational, and standard compliance answers for the Laboratory Cement Autoclave.

01What is the testing purpose of the Laboratory Cement Autoclave?

The Laboratory Cement Autoclave conducts accelerated soundness testing of Portland cement by detecting potential delayed expansion caused by unhydrated magnesia (MgO) or free lime under high steam pressure.

02Which standards govern the cement autoclave test?

It conforms strictly to IS 4031 (Part 3) / ASTM C151, specifically IS: 4031 (Part 3), IS: 269, and ASTM C151 specifications.

03What are the operating pressure and chamber construction details?

Features a Heavy-duty Stainless Steel (SS 304) rated for working steam pressure up to 21 kg/cm² (300 psi) and 215°C, complete with certified pressure gauge, safety blow-off valve, and rupture disc.

04What is the standard test cycle procedure?

Place neat cement prism bars on the interior specimen rack with water. Heat until steam pressure reaches 21 kg/cm² within 45–75 minutes, hold for 3 hours, cool down, and measure linear expansion in a length comparator.

05What safety and maintenance checks are essential?

Inspect the silicone lid gasket and safety relief valve before each run. Ensure distilled water is used to prevent internal scale accumulation on heating elements.

06What warranty and safety certifications are included?

Supplied with hydrostatic factory pressure test certificates and a 12-month manufacturer warranty from Maruti Sagar Engineering.

Need accessories?

We stock tripods, leveling staffs, and protective cases for all our surveying instruments.

Contact Sales